Programming guidelines 47
3. Programming guidelines
Only the Supervisor is allowed to program the menus, Electrode,
Reagent, and Method.
To select the Supervisor mode, press Stop for 3 seconds in the Main
window and enter the Supervisor code. Press 5 (Exit) and select
Return in mode = Supervisor.
* IMPORTANT *
For rst time users, it is recommended to program the instrument as
follows:
1. Create the electrode(s) to be used to perform the measurements.
2. Create the reagent(s) to be used during the titration.
Then nally,
3. Create the method(s), which will consequently use the electrode(s)
and reagent(s) created in the rst two steps of programming.
Once you have nished programming, make sure that NO question
marks "?" are displayed in the Reagent and Electrode tabs!