2.
TCM3
TCM3
and
TCM30
Service
Manual
Checkout
and
Adjustment,
Continued
Analog
Outputs
TCM3
from
RUN
22:
(continued)
Step
0:16
Step
0:17
Step
0:18
Max. Temp.
Watchdog
Step
0:19
Step
Action
1.
All
channels,
0%
modulation
Check
that
the
output
from
all
4
channels
(pO,
pCO,
3
and
4)
are
—109.0
+
5.0
mV.
If
outside,
adjust
the
pO;
channel
(by
R48
on
the
Power
Supply
PCB)
to
—109.0
+
1.0
mV.
Check
that
the
other
3
channels
are
—109.0
+
5.0
mV.
2.
All
channels,
20.12%
modulation
Check
that
the
output
from
all
4
channels
(pO.
pCO»,
3
and
4)
are
1000.0
+
15.0
mV.
K
outside,
adjust
the
pO,
channel
by
R67
on
the
Power
Supply
PCB
to
1000.0
+
1.0
mV.
Check
that
the
other
3
channels
are
1000.0
+
15.0
mV.
:
3
All
channels,
1.95%
modulation
Check
that
the
output
from
all
4
channels
(pO;,
pCO,,
3
and
4)
are
0.0
+ 5.0
mV.
If
the
tolerances
in
step
0:16,
0:17,
or
0:18
are
exceeded
after the
above
adjustments
have
been
performed,
the
Power
Supply
PCB
is
defective
and
must
be
replaced.
This
procedure
checks
the
Max.
Temp.
Watchdog.
In
this
step
the
CPU
does
not
trigger
the
watchdog
for
which
reason
the
heat
is
turned
off.
Step
1.
Action
Simulator
TEMPERATURE
37
°C.
Connect
DVM-
to
simulator
HEAT+
and
DVM+
to
GND.
Check
that
the
voltage
is
less
than
0.050
V.
Continued
on
next
page