OPERATION OF ACCESSORIES
- 32 -
4. VOC Zeroing Tubes
VOC Zeroing Tubes (P/N 025-2000-010) are used for
filtering organic air contaminants that may affect the
zero calibration reading.
1) Connect the tube adapter to the Flex-I-Probe by inserting
the tip of the sample probe into the smaller end;
2) Break the two ends of a VOC zeroing tube using the side
hole of the tube adapter;
3) Insert one end (the arrow in the silk screen indicates the
right direction) of the open VOC zeroing tube into the bigger
end of the adapter;
4) Run the zeroing calibration procedure of the instrument to
zero the unit;
5) Discard the used VOC zeroing tube (single use only) and
put the tube adapter back to the VOC zeroing kit container.
Tube tip breaker
Tube adapter
VOC Zeroing Tube
Flex-i-Probe