CustomAir by RAMVAC compressors offer superior air quality, quiet operation,
minimal space requirements and compressor heads that can be isolated for service.
CustomAir compressor heads are 100% duty cycle rated allowing them to run
continuously if needed.
Dual column desiccant dryer provides 100% continuous supply of moisture free air.
Our solid state electronic controller utilizes state-of-the-art circuitry and monitors
system functions which allows for multiple compressors to be operated
simultaneously in larger clinics.
CustomAir compressors are designed using trouble-free solid state transducers, in
lieu of mechanical pressure switches, to ensure dependability and long-life.
Like RAMVAC's dependable vacuum systems, CustomAir compressors utilize Baldor
motors for years of service.
CustomAir by RAMVAC Compressors have the best warranty in the industry -
6 years/4200 hours!
UL60601 Listed, NFPA 99c Compliant
For any practice looking for less
time servicing their compressor,
CustomAir® by RAMVAC®
is the number one choice.
M A K E S Y O U R P R A C T I C E P E R F E C T ®
CustomAir® by RAMVAC® Compressors
www.DentalEZ.com
http://Blog.DentalEZ.com

Questions and Answers

  • K
    Kevin TranAug 12, 2025
    Why RAMVAC Dental equipment compressor runs but pressure does not increase?
    • V
      Victoria RochaAug 12, 2025
      If your RAMVAC Dental equipment compressor runs but the pressure isn't increasing, consider the following: * Leaks in compressor fittings: See “Test Compressor for Leaks”. * Leaks in facility air lines: Test facility air lines for leaks. * Dirty intake filters: Replace the intake filters.
  • S
    Stephanie GarciaAug 18, 2025
    Why does my RAMVAC CustomAir 1000 Dental equipment compressor start and shut off more frequently than normal?
    • F
      felicia71Aug 18, 2025
      If your RAMVAC Dental equipment compressor starts and shuts off more frequently than normal, consider the following causes: * Condensate in tank: Contact RAMVAC®. * Leaks in compressor fittings: See “Test Compressor for Leaks”. * Leaks in facility air lines: Test facility air lines for leaks.
  • F
    fishermaryAug 27, 2025
    Why does RAMVAC CustomAir 1000 compressor run when pressure is not being used?
    • L
      Luis FaulknerAug 27, 2025
      If your RAMVAC Dental equipment compressor is running when pressure is not being used, there might be: * Leaks in compressor fittings: See “Test Compressor for Leaks”. * Leaks in facility air lines: Test facility air lines for leaks.
  • R
    Robert JohnsSep 3, 2025
    How to cool down a RAMVAC CustomAir 1000 Dental equipment compressor that gets too hot?
    • R
      Randy JenkinsSep 3, 2025
      If your RAMVAC Dental equipment compressor gets too hot, here are a couple of things to check: 1. **Ambient Temperature**: Ensure the ambient temperature is not exceeding 40ºC/ 104ºF. Lower the ambient temperature if it is too high. 2. **Compressor Operation**: Refer to the troubleshooting steps for “Compressor runs but pressure does not increase.”, as the compressor might not be shutting off properly.