Trying Out the Instrument
R&S
®
RTM2000
35Getting Started 1317.4710.02 ─ 05
1. Press the CURSOR key.
The cursor lines are switched on.
2. Press "Meas. Type" until "Voltage & Time" is highlighted.
3. Press "Source" until "CH1" is highlighted.
4. Press the NAVIGATION rotary knob to select the first cursor. Turn the rotary knob
to move the cursor line to the beginning of a pulse.
5. Press the NAVIGATION rotary knob again to select the second cursor. Turn the
rotary knob to move the cursor line to the end of the pulse.
6. Repeat the step to position the horizontal cursors.
The following results are displayed in the result table.
Δt: pulse width
t1, t2: time of the cursor positions, measured from the trigger point
ΔV: difference between top and base level
V1: top level voltage
V2: base level voltage
Performing Cursor Measurements