EasyManua.ls Logo

RaySafe Solo - Time

RaySafe Solo
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 4. kV and kVp
Table: Active compensation on kV measurements
SENSOR FILTRATION kV/KVp
R/F low Max. 15 mm Al kV
R/F high RAD Max. 15 mm Al kVp (when signal level is high enough)
DENT low Max. 1 mm Cu or equivalent kV
DENT high Max. 1 mm Cu or equivalent kVp (when signal level is high enough)
Use a kVp delay on machines with a slow rising output. 150 ms on dental single phase machines and 1000 ms
on a fluoroscopy machine is recommended.
kV is available for R/F, MAM, RAD and DENT detectors.
TIME
The RaySafe Solo measures time from the start trig (may be adjusted with the Trig level setting) until the signal
falls below 25 % of the peak (50 or 75 % if adjusted with the Trig level setting). At low dose rates (about 1 %
of the max dose rate for the active sensor), the 25 % end level is changed to a low level (about the lowest
measurable dose rate for the active sensor).
If the radiation has a pulsed characteristic, the time is measured until the last pulse ends. The dead time
interval between pulses must, however, be less than the Calc. Delay time (0.5, 2, 4, 6 or 7 s). The R/F low
or DENT low sensor has an electrical bandwidth of 0.1 kHz (slower rising and falling slopes), causing the
displayed exposure time to be a few milliseconds longer than for the R/F high or DENT high sensor, which has a
bandwidth of 2.5kHz.
29
RaySafe Solo User Manual – Measurement definitions