TIPS AND TRICKS
ARRANGEMENT OF MEASUREMENT
Make sure nothing is in the X-ray field when measuring with the RaySafe Solo.
The fact that the dose is proportional to 1/d
2
(where d is the FDD – Focus to detector Distance) may be used
when positioning the sensor.
d
SENSOR POSITION
The solid state sensors in the RaySafe Solo detector are positioned approximately 7 mm beneath the detector
surface. The position is marked with a line on the aluminium housing. This position is used as a calibration
reference point for dose.
RESET SWITCH
If the software hangs up, by any reason, activate the system reset by pressing a thin object through the small
hole beside the RS-232 connector on the side of the base unit.
If the detector hangs up when the battery is down and does not respond even when a charger is connected,
reset as described above and charge for some minutes before turning the instrument on again.