Calibration Instructions
1. Fillacontainerwithcrushedice.
2. Add water until ice is completely submerged.
3.
Add more ice to ensure it is packed tightly at the bottom of the container.
4.
Insertthethermometerstemandallowtostabilizeforapproximately3-5
minutes, while not allowing the tip to touch the bottom of the container
5. Iftheunitdoesnotmeasure32ºF(0°C)oncethereadinghas
stabilized, slowly turn the calibration screw using a at head
screwdriveruntilthemetermeasures32ºF(0°C).
Note: Turn the calibration screw to the left to decrease the temperature
value or to the right to increase the temperature value.
Battery Replacement
1. Openthebatterydoorbyslidingaatheadscrewdriverunderthelip
of the rubber battery door.
2.
When open, tap the unit against your hand to make the battery slide out.
3. InsertanewLR44battery.
4. Push the rubber door in rmly so that the battery is all the way
in position.
Accessories and Replacement Parts
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.
Product Care
To keep your instrument in good working order we recommend the following:
• Storeyourproductinaclean,dryplace.
• Changethebatteryasneeded.
• Ifyourinstrumentisn'tbeingusedforaperiodofonemonthorlonger
please remove the battery.
• Cleanyourproductandaccessorieswithbiodegradablecleaner.Donot
spraythecleanerdirectlyontheinstrument.Useonexternalpartsonly.