H4i™
HEATED HUMIDIFIER
User Guide
English
Global leaders in sleep and respiratory medicine www.resmed.com

Other manuals for ResMed H4i

Question and Answer IconNeed help?

Do you have a question about the ResMed H4i and is the answer not in the manual?

ResMed H4i Specifications

General IconGeneral
TypeHeated humidifier
Water Chamber Capacity380 ml
Humidification TypeHeated
CompatibilityResMed S8 series CPAP machines
Power SupplyAC

Summary

Introduction to H4i Humidifier

Indications for use

Specifies the intended use of the H4i for humidifying air from ResMed compatible devices.

Contraindications

Lists conditions where the H4i should not be used, especially with bypassed upper airways.

Components of your H4i

Details the parts of the H4i humidifier, including the cover, chamber, and docking station.

Setting up your H4i for use

Provides initial steps for preparing the H4i for connection to a flow generator.

System Setup and Connection

System setup

Instructions for physically connecting and setting up the H4i with the flow generator and power.

Physical connection and power

Steps for aligning the H4i, connecting the power cord, and plugging into the outlet.

Air tubing connection

Steps for attaching the air tubing to the H4i's air outlet and mask system.

Safety warnings for connection

Critical safety advice including powering off, chamber status, and preventing water ingress.

Invasive use contraindication

Notes contraindication for invasive use and recommends approved external humidifiers.

Filling and Detaching the Water Chamber

Filling the water chamber

Step-by-step guide on how to fill the H4i water chamber with clean water.

Open and remove water chamber

Instructions for opening the H4i cover and detaching the water chamber for filling.

Fill with fresh, clean water

Guidance on filling the water chamber up to the maximum water level mark.

Caution: Overfilling and additives

Warnings against overfilling and using additives which can affect performance or damage materials.

Detaching your H4i

Procedure for safely disconnecting and removing the H4i from the flow generator.

Unplug, cool, and detach H4i

Steps to detach the H4i, including unplugging, cooling hot parts, and releasing the unit.

Starting and Stopping Treatment

Warming H4i with VPAP/S8 or Stellar 150

Instructions for using the Warm-Up feature on different ResMed device models.

Caution: Pressure limits

Important caution about maximum treatment pressure limits for the H4i.

Starting your treatment

Steps to begin a therapy session, including adjusting humidity and positioning tubing.

Adjust humidity and position tubing

Guidance on adjusting the control dial for humidity and arranging tubing for comfort.

Maintenance and Disassembly

Warning: Excessive air tubing length

Safety warning about excessive tubing length to prevent entanglement during sleep.

Stopping your treatment

Instructions on how to cease a therapy session.

Cleaning and maintenance

Essential procedures for maintaining the H4i, water chamber, and accessories.

Warnings for cleaning

Critical safety advice for cleaning, covering electrical hazards and hot components.

Disassembling the H4i cleanable water chamber

Step-by-step guide for taking apart the cleanable water chamber for cleaning.

Prepare chamber and disengage latch

Steps for removing spill guard, water, and disengaging the latch mechanism.

Remove tub plate and seal

Steps for removing the tub plate and its seal from the water chamber.

Water Chamber Cleaning Procedures

Cleaning your H4i Water chamber

Daily and weekly cleaning procedures for the water chamber, spill guard, and related parts.

Daily water chamber cleaning

Daily cleaning procedure for the water chamber, spill guard, tub plate, and seal.

Weekly water chamber cleaning

Weekly cleaning advice, including dishwasher use and inspection for damage.

Cleaning water chamber deposits

Method for cleaning deposits using a vinegar and water solution.

Cover, Docking Station, and Reassembly

Cover and cover seal maintenance

Procedures for removing, cleaning, and inspecting the H4i cover and its seal.

Remove and inspect cover/seal

Instructions for detaching the cover, peeling the seal, and inspecting for wear.

Docking station and heater plate care

Maintenance procedures for the docking station and heater plate.

Periodically wipe docking station

Periodic cleaning of the docking station's exterior with a damp cloth.

Caution: Prohibited cleaning agents

Warning listing cleaning agents that can damage the H4i and reduce its lifespan.

Keep heater plate dry

Instruction to keep the heater plate dry and wipe with a clean cloth if necessary.

Reassembling the H4i cleanable water chamber

Step-by-step guide for reassembling the H4i cleanable water chamber.

Reassemble seal, plate, and catch

Steps for refitting the tub plate seal, reinserting the tub plate, and locking the catch.

Reassembling and Servicing

Reassembling your H4i

Steps for putting the main H4i unit back together after cleaning.

Align cover seal and attach cover

Steps to align the cover seal and attach the cover to the docking station.

Replace water chamber

Final step to place the water chamber back into the humidifier docking station.

Servicing requirements

Information on when and how the H4i should be serviced by an authorized agent.

Frequently asked questions

Answers to common user queries regarding condensation, water, and dryness.

Common user questions

Addresses condensation, use of distilled water, and symptoms of dryness.

Troubleshooting Guide

Troubleshooting common issues

Guide to resolving problems like device not working, low airflow, or leaks.

Device not working or no indicator light

Troubleshooting steps for a non-functional device, checking power and connections.

Indicator light dull or no heat

Diagnoses issues with the indicator light or lack of heat, suggesting cleaning.

Leaking water chamber

Troubleshooting for leaks, recommending checking chamber condition and tub plate fit.

No or low air flow

Addresses issues with air flow, checking tubing connection, blockages, and generator status.

Water splashing on face

Solutions for water splashing, involving overfilling and condensation management.

Technical Specifications and Performance

Technical specifications

Detailed specifications for the H4i, including temperature, dimensions, and power.

Temperature and safety specs

Specifies heater plate temperature, safety cut-out, and maximum delivered gas temperature.

Dimensions, weight, and water capacity

Provides physical size, weight, and maximum water volume for the H4i components.

Power supply and materials

Details power requirements and materials used for the docking station and water chamber.

Operating and storage conditions

Defines acceptable temperature and humidity ranges for operation and storage.

IEC classification and EMC

Indicates safety standards compliance and refers to EMC information.

Humidifier performance

Presents data on humidifier performance at specified ambient conditions.

Symbols, Environment, and General Warnings

Symbols which appear on the device

Explanation of various symbols found on the H4i device and its packaging.

General safety and handling symbols

Explains common safety symbols like caution, reading instructions, and handling precautions.

Device and component identification symbols

Explains symbols for European Representative, applied parts, and equipment classification.

Control and indicator symbols

Explains symbols for humidifier controls and maximum water level indicators.

Environmental information

Information on the European directive for proper disposal of electrical and electronic equipment.

WEEE Disposal Guidelines

Information on the European directive for proper disposal of electrical and electronic equipment.

General warnings

Important safety warnings and precautions for using the H4i humidifier.

Intended use, compatibility, accessories

Warnings on intended use, device compatibility, and using only ResMed-recommended accessories.

Positioning, condensate, patient risks

Discusses risks related to device positioning, condensate, and potential patient harm.

Service, spills, damage, electrical safety

Precautions regarding servicing, liquid spills, device damage, and electrical safety.

Supplemental oxygen precautions

Reminder to adhere to safety precautions when supplemental oxygen is in use.

Limited Warranty Information

Limited warranty

Details the warranty period, exclusions, and claims process for ResMed products.

Product Warranty Periods

Details warranty periods for various ResMed products including the H4i.

Warranty Exclusions and Limitations

Lists conditions and damages not covered by the ResMed limited warranty and claim limitations.