Problem/possible cause Solution
The following message is NOT displayed on the LCD after you try to update the settings using
the SD card: Settings updated successfully, press any key
The settings were not updated.
Contact your clinician/service provider immediately.
General technical specifications
Power supply
90W power supply unit
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
30W power supply unit
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 20W (30VA)
Maximum power consumption: 36W (75VA)
Nominal inputs: 12V, 24V
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
Environmental
conditions
Operating temperature: +5°C to +35°C
Operating humidity: 10 to 95% non-condensing
Operating altitude: Sea level to 2,591m
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: 10 to 95% non-condensing
Aircraft use ResMed confirms that the device/s meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160F, section 21, category M) for all phases of air
travel.
Electromagnetic
compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. Information regarding the electromagnetic emissions and
immunity of this ResMed device can be found on www.resmed.com, on the
Products page under Service and Support. Click on the PDF file for your
language.
IEC 60601-1
classification
Class II (double insulation), Type BF