7
English
Power level (CPAP mode)
With SlimLine tube: 34 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
With Standard tube: 35 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
With either SlimLine
or Standard tube and
H5i:
36 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
Supplemental
oxygen
Recommended maximum supplemental oxygen flow: 15 L/min
Air tubing technical specifications
ClimateLine SlimLine
ClimateLine
MAX
Standard 3 m
Maximum
recommended
pressure
25 cm H
2
O 25 cm H
2
O 25 cm H
2
O 25 cm H
2
O 25 cm H
2
O
Inner diameter
Length 1.98 m 1.8 m 1.91 m 2.0 m 3.0 m
Material Flexible
plastic and
electrical
components
Flexible plastic Flexible plastic
and electrical
components
Flexible
plastic
Flexible
plastic
Heated tubing
temperature cut-
out
≤ 41°C
≤ 41°C
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• The temperature and relative humidity settings displayed for ClimateLine / ClimateLine
MAX
are not
measured values.
• Check with your clinician/service provider before using the SlimLine air tubing with devices other
than the S9 or H5i.
• The electrical connector end of the heated air tubing is only compatible with the H5i air outlet and
should not be fitted to the device or mask.