57
AUTORUN
CHECK
It appears when middle ear
analyzer tests are performed
when the Autorun function has
been activated
System
message
It means that the test is
being performed with
the Autorun function
activated
No action is required–if
the Autorun function
is not wanted, it can be
deactivated via the Setup
menu
DO YOU
CONFIRM ?
This query appears after entry of
plot cancellation or parameter
change requests
System
message
It means that a plot
cancellation or a
parameter change
request has been
entered
Press either the YES or
the NO when the query
appears
CHANGE
CALIBRATION
DATE ?
This message appears at the end
of the device calibration proce-
dure, once the “SAVE” softkey has
been pressed and values have
been changed.
System
message
It provides the option
of updating periodic
calibration dates
Press either the YES or
the NO when the query
appears
INSERT PROBE
INTO THE
SPECIFIED CAVITY
This message appears when
on the Compliance Calibration
page
System
message
This message refers
to the cavity value
displayed on the Cali-
bration page, that it is
necessary to insert the
probe into for calibra-
tion performance
Insert the probe into the
test cavity as displayed
PRESS
CONFIRM/AUTO-
MATIC
KEY
This message comes up on the
Compliance Calibration page
the instant in which the probe is
inserted into the specied cavity
System
message
It provides instruc-
tions that need to be
followed in order to
proceed with compli-
ance calibration
Press the conrmation
key to calibrate selected
probe tone frequency
or automatic key to
calibrate all probe tone
frequencies (PT models)
CHECK CAVITY This message is given on the
Compliance Calibration page if
the probe is not inserted into the
displayed cavity
System
message
It means that the probe
has been inserted into
the wrong cavity
Proceed with inserting
the probe into the
correct test cavity, as
displayed
READ
VALUE (ml)
This message is given on the
Compliance Calibration page, at
the end of the calibration process
System
message
It means that the value
that is displayed is the
real-time value reading
by the device
No action is required
INSERT PROBE
INTO THE 2cc
CAVITY
This message appears on the
ARLT Calibration page, if the
probe is not inserted into the
2cc cavity.
System
message
You are instructed to
insert the probe into
the 2cc cavity in order
to perform calibration
Proceed with inserting
the probe into the 2cc
cavity
PRESS RUN TO
PERFORM
CALIBRATION
This message appears on the
ARLT Calibration page, when
the probe is inserted into the
2cc cavity
System
message
It provides instructions
on proceeding with
calibration
Press the Start key as
instructed
ENTER
PASSWORD
This message appears when
attempting rst-time access to
the device calibration procedure,
once the unit has been switched
on
System
message
It instructs you to enter
your access password
to unlock the device
calibration procedure
Enter your password via
the softkeys
WRONG
PASSWORD
This message appears when
attempting to access the device
calibration procedure using a
wrong password
System
message
It means that you have
entered the wrong
password for access to
the Calibration menus
Press the Enter key, then
re-enter the password
correctly

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Resonance r25c and is the answer not in the manual?

Resonance r25c Specifications

General IconGeneral
BrandResonance
Modelr25c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals