EasyManuals Logo

Resonance r25c User Manual

Resonance r25c
Go to English
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
61
Reex, Decay and ARLT tests
Tone frequency accuracy: ±1%
dB level variation: 1, 2, 5 or 10 dB
Stimulus duration (sec):
Reex: 0.5, 1 or 2
Decay: 10
ARLT: 1
Non acoustic: 10 or 20
Max frequencies and max levels
Tone frequency accuracy: ±1%
dB level variation: 1, 2, 5 dB
Audiometry section
Tests
Pure tone Audiometry and Speech
Audiometry (speech material recorded
on SD card)
ABLB; MLB; DLI; SISI
Tone Decay, Multifrequency, Stenger
Automatic threshold (Autothreshold)
Audiometry, Békésy
GAP, DLF, Quick SIN (optional)
Outputs
ACR, ACL, BC, INSERT, Free Field, monitor
phone
Inputs
CD tape (Ext1 and Ext2), mike talk over
Type of signals
pure tone, warble, mike, external, speech
material recorded on SD card
Masking
Narrow band noise, speech noise, white
noise, external signal.
DD45 - TDH39 HDA280 INSERT BC B71 (dBuN) Free Field
freq.
(Hz)
Tones
(dBHL)
Noise
bands
(dBEM)
Tones
(dBHL)
Noise
bands
(dBEM)
Noise
bands
(dBEM)
Tones
Noise
bands
Tones
(dBHL)
Noise
bands
(dBEM)
125 80 55 80 55 ---- ---- ---- 70 70
250 100 75 100 75 75 45 40 85 85
500 120 90 120 90 95 70 60 95 95
750 120 95 120 95 95 75 65 95 95
1000 120 95 120 95 95 80 70 95 95
1500 120 95 120 95 95 80 75 95 95
2000 120 95 120 95 95 80 75 95 95
3000 120 95 120 95 95 80 70 95 95
4000 120 95 120 95 95 80 70 95 95
6000 115 90 110 90 80 60 55 95 95
8000 100 70 100 70 80 55 50 85 85
10000 ---- ---- 90 70 ---- ---- ---- ---- ----
12500 ---- ---- 80 70 ---- ---- ---- ---- ----
WN 110 (dBspl) 110 (dBspl)
110
(dBspl)
100 95 (dBspl)
SN 90 90 90 60 95
EXT 105 105 ---- 60 90
DLI and SISI Tests
Increment levels SISI: 0.25, 0.5, 0.75, 1, 1.5, 2, 3, 4, 5 dB
SISI increment recurrence rates: 0.2, 0.5 Hz, random
Increment levels DLI: 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1, 1.5, 2, 3, 4, 5 dB
DLI increment recurrence rates: 0.5, 1, 2 Hz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Resonance r25c and is the answer not in the manual?

Resonance r25c Specifications

General IconGeneral
BrandResonance
Modelr25c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals