66
Producer’s guide and declaration - Electromagnetic IMMUNITY
The R25C/R35C is suitable for use in the specied electromagnetic environment. The purchaser
or user of the The R25C/R35C should assure that it is used in an electromagnetic environment as
described below:
Immunity test IEC 60601-1-2
Test level
Compliance
Level
Electromagnetic
Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
2/4/8/15 kV air
IEC 60601-1-2
Test level
Floors should be wood. con-
crete or ceramic tile. If oors
are covered with synthetic
material. the relative humidity
should be at least 30%.
Radiated
electromagnetic
eld
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2
Test level
Portable and mobile RF
communications equipment
should be used no closer to
any part of the R25C/R35C
including cables.
Recommended istance:
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
Where P is the aximum power
output of the transmitter
in watt (W) based on the
transmitter’s producer and d is
the recommended distance in
meters (m).
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines > 3 m
IEC 60601-1-2
Test level
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
0.5/1 kV dierential
mode
0.5/1/2 kV common
mode
IEC 60601-1-2
Test level
Mains power quality should be
that of a typical commercial or
hospital environment.
Conducted
disturbances
induced by RF elds
IEC 61000-4-6
3 V
150 kHz to 80 MHz
IEC 60601-1-2
Test level
Portable and mobile RF com-
munications equipment
should be used no closer to
any part of the R25C/ R35C in-
cluding cables.
Recommended distance d =
1,2√P