43
INSTRUCT
PATIENT
It appears during performance of
the ETF testing for non-perfora-
ted eardrums
System
message
It means that the
patient is required
to swallow, when
instructed by the
operator to do so
No action is required
PROBE NOT
CONNECTED
It appears during middle ear
analyzer testing, if the probe is
not connected up to the device
Error
message
It means that the probe
is not hooked up to the
appropriate connector
on the device’s rear
panel
Connect the probe up
to the correct rear-panel
connector
AUTORUN
CHECK
It appears when middle ear
analyzer tests are performed
when the Autorun function has
been activated
System
message
It means that the test is
being performed with
the Autorun function
activated
No action is required - if
the Autorun function
is not wanted, it can be
deactivated via the Setup
menu
DO YOU
CONFIRM ?
This query appears after entry of
plot cancellation or parameter
change requests
System
message
It means that a plot
cancellation or a para-
meter change request
has been entered
Press either the YES or
the NO when the query
appears
CHANGE
CALIBRATION
DATE ?
This message appears at the end
of the device calibration proce-
dure, once the “SAVE” softkey has
been pressed and values have
been changed.
System
message
It provides the option
of updating periodic
calibration dates
Press either the YES or
the NO when the query
appears
INSERT PROBE
INTO THE
SPECIFIED CAVITY
This message appears when on
the Compliance Calibration page
System
message
This message refers to
the cavity value displa-
yed on the Calibration
page, that it is necessa-
ry to insert the probe
into for calibration
performance
Insert the probe into the
test cavity as displayed
PRESS
CONFIRM/AUTO-
MATIC
KEY
This message comes up on the
Compliance Calibration page
the instant in which the probe is
inserted into the specied cavity
System
message
it provides instructions
that need to be follo-
wed in order to pro-
ceed with compliance
calibration
Press the conrmation
key to calibrate selected
probe tone frequency
or automatic key to
calibrate all probe tone
frequencies (PT models)
CHECK CAVITY This message is given on the
Compliance Calibration page if
the probe is not inserted into the
displayed cavity
System
message
It means that the probe
has been inserted into
the wrong cavity
Proceed with inserting
the probe into the
correct test cavity, as
displayed
READ
VALUE (ml)
This message is given on the
Compliance Calibration page, at
the end of the calibration process
System
message
It means that the value
that is displayed is the
real-time value reading
by the device
No action is required
INSERT PROBE
INTO THE 2cc
CAVITY
This message appears on the
ARLT Calibration page, if the
probe is not inserted into the
2cc cavity.
System
message
You are instructed to
insert the probe into
the 2cc cavity in order
to perform calibration
Proceed with inserting
the probe into the 2cc
cavity
PRESS RUN TO
PERFORM
CALIBRATION
This message appears on the
ARLT Calibration page, when
the probe is inserted into the
2cc cavity
System
message
It provides instructions
on proceeding with
calibration
Press the Start key as
instructed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Resonance R26M and is the answer not in the manual?

Resonance R26M Specifications

General IconGeneral
BrandResonance
ModelR26M
CategoryMeasuring Instruments
LanguageEnglish