EasyManua.ls Logo

Riester ri-focus LED - Page 27

Riester ri-focus LED
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
Guidelines and manufacturer‘s declaration – electromagnetic immunity
The ri-focus LED headlights and clar N LED head mirrors are intended for use in the electromagnetic environment specified below. The customer or user of the
headlight should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Con: ±8 kV
Air: ± 2, 4, 8,
15 kV
Con: ±8 kV
Air: ± 2, 4, 8,
15 kV
Floors should be wood, concrete or ceramic tile. If the floor is covered
with synthetic material, the relative humidity must be at least 30%.
Rapid transients electri-
cal disturbances/bursts
IEC 61000-4-4
Not applicable Not applicable The quality of the supply voltage should be that of a typical commercial
or hospital environment.
Impulse voltage
IEC 61000-4-5
Not applicable Not applicable The quality of the supply voltage should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations according to
IEC 61000-4-11
Not applicable Not applicable The quality of the supply voltage should be that of a typical com-
mercial or hospital environment.
Magnetic field with
energy-efficient rated
frequencies
IEC 61000-4-8
30A/m 30A/m Magnet fields at mains frequency should be at a level typical for the
location of a typical commercial hospital environment.
Note: UT is the AC source. Mains voltage before the application of the test level.

Table of Contents