Riester ri-sonic PCP-USB - User Manual

Riester ri-sonic PCP-USB
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
Instruction of use
Electronic Stethoscope
ri-sonic PCP-USB
ri-sonic PCP-1
Question and Answer IconNeed help?

Do you have a question about the Riester ri-sonic PCP-USB and is the answer not in the manual?

Overview

The Riester ri-sonic PCP-USB and ri-sonic PCP-1 are electronic stethoscopes designed for converting internal body sounds, such as heart and lung sounds, into electronic signals for delivery to a compatible PC or mobile device. These devices are manufactured according to Directive 93/42 EEC and are intended for use by suitably trained personnel, such as doctors or medically trained persons (e.g., nurses), in a professional healthcare facility environment.

Function Description

The electronic stethoscope consists of two main elements: a chest piece and a cable with a plug.

  • ri-sonic PCP-USB: This model features a chest piece connected via a straight cable with a USB plug. The chest piece contains an embedded piezo sensor, an audio amplifier, an Analog to Digital Converter (ADC), and an Encoder to create a digitized stream. This data is then sent to the PC via the USB interface. The device derives its operating voltage from the 5V lead of the USB interface.
  • ri-sonic PCP-1: This model features a chest piece connected via a spiral cable with a 3.5mm stereo plug. The chest piece assembly includes a Printed Circuit Board Assembly (PCBA), a piezo element (PE), an Enclosure Top piece, an Enclosure Bottom Piece, and a cable. The PCBA provides amplifier circuitry to raise the signal level from the PE to a level comparable to an electret microphone. This amplifier requires an operating voltage between 2 VDC and 5 VDC, typically supplied by the microphone input port of the connected device.

Both models are designed to provide sound data as specified, and the user must verify that the sensor's line data matches the requirements of the PC, mobile device, app, or software being used.

Important Technical Specifications

  • Mode of operation: Continuous operation.
  • Electrical protection: Class II protection against electrical shock; Type BF applied part.
  • Voltage Input: 2V-5V DC (from PC).
  • Auscultation bandwidth (Frequency range): 20 Hz to 2,000 Hz.
  • Weight:
    • ri-sonic PCP-USB with cable: 180g
    • ri-sonic PCP-1 with cable: 145g
  • Operating conditions: 5°C to 40°C at a water vapor pressure up to a relative humidity range of 30% to 75%.
  • Storage and transport conditions: -25°C to 35°C (up to 70°C for short periods) at a water vapor pressure up to 50mbar and a relative humidity range of 0% to 90%.
  • Air pressure: 700hPa to 1060hPa.
  • System requirements:
    • PC with Windows 10
    • Apple Computer with Apple macOS 10.13.6
    • Audio support (sound card or onboard sound)
    • USB or Microphone port
    • Speaker Frequency coverage (20 – 2000 Hz)
  • Electromagnetic Compatibility (EMC): The devices are Group 1, Class B according to CISPR 11, meaning they use RF energy only for internal functions and have very low RF emissions. They are intended for use in all establishments, including residential areas.
  • Defibrillation-proof: This product is not a defibrillation-proof applied part.
  • Anesthetic mixture compatibility: Not suitable for use in the presence of a flammable anesthetic mixture with air or with continuous oxygen or nitrous oxide.

Usage Features

  • Installation:
    • ri-sonic PCP-USB: Simply plug the USB connector into the USB port of the PC. No additional software is required for basic PC usage. For specialized software, follow the software's instructions.
    • ri-sonic PCP-1: Plug the 3.5mm stereo plug into the Microphone port of the PC. No additional software is required for basic PC usage. For specialized software, follow the software's instructions.
  • Auscultation Sessions: The stethoscope is connected to a PC before an auscultation session. The chest piece is positioned on the patient's body (upper body) to capture body sounds, allowing doctors to hear them and satisfy clinical objectives.
  • Patient Population: Intended for all patients.
  • Environmental Conditions: Designed for use in a Professional Healthcare Facility Environment. If exposed to environmental conditions outside specified ranges, the unit should be checked for proper operation before reuse.
  • Cables: Cables are a strangulation hazard; keep them out of reach of infants and small children.

Maintenance Features

  • Cleaning and Disinfection:
    • To prevent cross-contamination, the stethoscope must be cleaned and disinfected regularly before each use.
    • The exterior can be cleaned with a damp cloth, optionally moistened with alcohol, until optically clean.
    • Use disinfection products (e.g., Disinfectant Bacillol AF from Bode Chemie GmbH / time 30s) strictly according to the manufacturer's instructions and national guidelines for proven effectiveness.
    • After disinfection, wipe the instruments with a damp cloth to remove any disinfectant residue.
    • Important: The ri-sonic PCP-USB / ri-sonic PCP-1 Chest Piece is not a sterile device and does not require sterilization. Do not immerse stethoscopes in liquids. They are not designed for machine-processed maintenance and sterilization, as this may cause irreparable damage.
  • Preventive Inspection/Maintenance/Calibration: The ri-sonic PCP-USB / ri-sonic PCP-1 Chest Piece does not require preventive inspection, routine maintenance, or calibration.
  • Troubleshooting (No stethoscope sounds from PC):
    • Test the local loop back through the PC Control Panel to check for a failed ri-sonic PCP-USB / ri-sonic PCP-1 Chest Piece.
    • Substitute the device with a replacement ri-sonic PCP-USB / ri-sonic PCP-1 Chest Piece to identify a failed unit.
  • Disposal: Batteries and electrical/electronic devices must be disposed of according to locally applicable regulations, not with domestic waste. Contact the manufacturer or representatives for disposal inquiries.
  • Warranty: A 2-year warranty from the date of purchase covers defects due to material or manufacturing faults. It does not cover improper handling or wearing parts. A valid warranty claim requires a completed and stamped warranty card from the dealer.

Riester ri-sonic PCP-USB Specifications

General IconGeneral
BrandRiester
Modelri-sonic PCP-USB
CategoryMedical Equipment
LanguageEnglish