User Manual for DM3000 Series
Ratio Measurement
Ratio measurement is used to measure the ratio of 2 directions DC voltage signal.
Ratio measurement is only for measuring DC voltage.
Press Æ Ratio , enter the menu shown below:
Figure 2-6
Table 2-7 Ratio Measurement Menu Explanation
Function Menu Explanation
ON Open the Ratio Measurement Function
OFF Close the Ratio Measurement Function
Save all changes, back to a higher level menu.
The method of the radio measurement:
Ratio=
tagefrenceVaolDC
VoltageDC
Re
(1). Measuring Sense Terminal, for measuring reference DC voltage. Default range
option is auto.
(2). Measuring Input Terminal, for measuring DC voltage. The measuring voltage
scope is 0V~1000V
(3). Input LO Terminal and Sense LO Terminal must have a common reference
value, and the voltage difference cannot surpass ±2V.
DC Volta
g
e
DC Reference Voltage