RIGOL Chapter 2 Functions of the Front Panel of GPSA
2-78 RSA3000 User Guide
Table 2-56 Gate Time
Parameter
Remarks
Default
100 ms
Range
1 μs to 500 ms
Unit
s, ms, μs, ns, ps
Knob Step
gate time/100, Min = 1 μs
Left/Right Arrow Key Step
Up/Down Arrow Key Step
at 1-1.5-2-3-5-7.5 step
Gate Time Auto
Turns on or off the auto gate time for marker counter.
Remarks:
Enables the auto gate time, and the gate time of the marker count is determined
by the instrument automatically.
When the auto gate time is disabled, press Gate Time to set it.