WORK INSTRUCTION M500-605-100-M01 Rev 1.00
Page 1 M500-605-100-M01
Sealing Methods C510, C520 and C527
Sealing with calibration seal stickers
Equipment required.
• Destructible stickers
• Setup switch cover
• Load cell connection cover (M5010)
Method
1. Place the setup switch cover in position so that it covers the setup switch and 2
closure screws.
2. Place a sticker over the centre of the setup switch cover and onto the side of the
enclosure as shown in Figure 1.