12
8. Voltage Measurement (V AC/DC)
You can use this instrument for voltage measurement from 25V to 600V
AC or DC.
* Keep the function selector switch to “V ” position and
apply the input.
* The instrument automatically displays whether the input
is AC or DC along with the value of the external voltage.
* If the value of voltage is < 25V (AC or DC), the display
“025V” flashes. If the value of the voltage is > 600V (AC
or DC), a continuos acoustic signal is generated.
* When reverse DC is applied it displays “-dc”
Additional features available in Insulation measurement ranges.
1. Minium and Maximum value “MIN/MAX” storage
Voltage Measurement
9. Default Live Circuit Warning
When more than 25V (AC or DC) is applied to the terminals the
instrument defaults to voltmeter on all switch positions. In addition the
beeper will sound on all switch position (except the V position).
When external voltage>25 V is present, Insulation Ohm and Continuity
Tests are inhibited.
When external voltage <25 V is present, only the Insulation Tests are
available !!
Note : The default live circuit warning does not operate
(i) when in STOPWATCH mode.
(ii) when in MENU structure.
+(-)/~
-(+)/~
2 92
MW
50V
100V
250V
500V
1KV
V
10 Stop Watch
This function allows you to measure elapsed time up to one hour.
To use stop-watch
1. Set the function selector switch to Continuity " " position.
2. Press “TEST” key. This will show the stopwatch from 00:00
3. Press key at any time to RESET the stopwatch.
4. Press key at any time to START or STOP the stopwatch.
5. To quit stopwatch and go back to basic display, Press “TEST” key.
11 Menu Structure
Structure of command menu for Insulation measurement :
Structure of command menu for Resistance & Continuity measurement
Basic
Display
nnnn
mm:ss
xxxx
OFF/ON
RISH
Insu
20
400
100kM
600
10kM
200
1MG
800
1kM
100k
1000
10MG
8
L
L
VAC VDC
Basic
Display
nnnn
xxxx
xxxx
OFF/ON
Basic
Display
OFF/ON
13
Structure of command menu for Voltate measurement
= Backlit off
= Backlit on
= Time
= Recall
= Minutes
= Seconds
= Value for noGo
= Strored value
mm
ss
nnnn
xxxx
OFF
ON
Esc
= Confirmation of
selected menu item
via ENTER key
= Select next menu item
in the flow direction
= Go back to previous
level of menu
CE