Electromagnetic Interference
Thisproductisdesignedandbuilttominimizeelectromagneticinterferencewithotherdevices.
However,ifinterferenceisnoticedbetweenanotherdeviceandthisproduct:
• Removeinterferingdevicefromroom.
• Plugtableintoanisolatedcircuit.
• Increaseseparationbetweentableandinterferingdevice.
• ContactMidmarkifinterferencepersists.
Disposal of Equipment
Attheendofproductlife,thetable,accessories,andotherconsumablegoodsmaybecomecontaminated
fromnormaluse.Consultlocalcodesandordinancesforproperdisposalofequipment,accessoriesand
otherconsumablegoods.
Transportation / Storage Conditions
Ambient Temperature Range:.....................................-5°C to +38°C (+23°F to 100°F)
Relative Humidity........................................................10% to 90% (non-condensing)
Atmospheric Pressure................................................500hPato1060hPa(0.49atmto1.05atm)
Intended Use
This product is intended to be used as a table to provide positioning and support of patients during
examinationproceduresconductedbyqualiedmedicalprofessionalsinanoceenvironment.
Electrical Requirements
Caution
To completely isolate the table from electrical mains supply, both power cords must be
unplugged.
When using high frequency surgical devices or endocardial catheters:
• Use non-conductive material to insulate patient from metal portions of the table.
• Consult operating instructions for the device before using in conjunction with drawer or
upholstery heater.
Failure to comply may result in electric shock or burns to patient.