Roche cobas c - Changing the Fuses

Roche cobas c
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
18 Operator’s Manual · Version 1.2
Maintenance cobas c pack opener
a Changing the fuses
1
Switch off the unit and disconnect the power cord.
If the power cord is still connected, you will not be able to access the fuse holder.
2
Using the tip of a blade screwdriver, lift the fuse holder out of the unit.
3
Replace both fuses.
Make sure you are using the correct types:
(Type 2A / 250V AC (5 x 20 mm) - Blow Characteristic -Time Delay).
4
Reconnect the power cord to the unit.
5
Switch on the unit.
Figure 8 Lifting the fuse holder