R&S FSH Power Meter
Using a Power Sensor
Operating Manual 1173.6275.12 - 30 138
Taking additional loss or gain into account
At high powers that cause the power sensor maximum input level to be exceeded or at
very low levels that are below the R&S FSH minimum sensitivity, the R&S FSH can
take additional loss or gain between the DUT and the power sensor into account.
These are defined in terms of an offset in dB relative to the measured level. A positive
offset corresponds to a loss and a negative offset to a gain.
The R&S FSH shows the current offset in the diagram header.
Press the AMPT key.
Press the "Ref Offset" softkey.
The value entry box for the reference offset opens.
Enter the required offset.
The offset is taken into account in the power or level display.