Performance Test R&S FSH
1145.5973.82 1.2 E-11
Performance Test
Checking the frequency accuracy
Test equipment: Signal generator (refer to section "Measurement Equipment and
Accessories", item 1)
Test setup: Connect the signal generator to the RF input of the R&S FSH.
Signal generator settings: - Frequency 1 GHz
- Level -30 dBm
R&S FSH settings:
- [ PRESET ]
- [ FREQ : 1 GHz ]
- [ SPAN : 100 kHz ]
- [ BW : MANUAL RES BW : 10 kHz ]
- [ MARKER : MARKER MODE : FREQ COUNT ]
Measurement: Read out the frequency value (Count:) of the marker.
Nominal frequency:............................................................1.0 GHz ± 1 kHz
Note:
The frequency of the reference oscillator can be adjusted by means of a
service function (refer to Chapter 2 "Adjustment").