R&S FSH Test Instructions
1145.5973.82 1.1 E-11
1 Performance Test
Test Instructions
• The rated specifications of the analyzer are tested after a warm-up time of at least 15 minutes. Only
by adhering to this requirement can compliance with the guaranteed data be ensured.
• Values specified in the following sections are not guaranteed. Only the technical specifications
provided on the data sheet are binding.
• The values specified in the data sheet are the guaranteed limits.
• Inputs for settings during measurements are shown as following:
[<KEY>] Press a key on the front panel, eg [SPAN]
[<SOFTKEY>] Press a softkey, e.g. [MARKER -> PEAK].
[<nn unit>] Enter a value and terminate by entering the unit, e.g. [12 kHz].
Successive entries are separated by [:], e.g. [ BW : MANUAL RES BW : 3 kHz ].
Measuring Equipment and Accessories
Item Type of
equipment
Specifications recommended Equipment
recommended
R&S order
no.
Use
1 Signal generator Frequency:
R&S°FSH3 10 MHz to 3 GHz
R&S°FSH6 10 MHz to 6 GHz
Uncertainty of frequency: 0.1 ppm
Phase noise at 500 MHz:
< -100 dBc (1Hz) @ 10 kHz
< -110 dBc (1Hz) @ 100 kHz
< -130 dBc (1Hz) @ 1 MHz
R&S SML03
R&S SMT06
1090.3000.13
1039.2000.06
Frequency response
Frequency accuracy of
reference oscillator
2 6-dB divider
(power splitter)
Weinschel
1870A
Frequency response
3 Power meter R&S NRP 1143.8500.02 Frequency response
4 Power sensor 1 MHz to 6 GHz
RSS 0.8%
Meter noise 20 pW
R&S NRP-Z91 1168.8004.02 Frequency response
5 N cable Attenuation < 1 dB to 6 GHz Tracking generator output
level
6
50-
termination
1 MHz to 6 GHz
Return loss -10 dB
Noise display