Measurement Equipment and Accessories for Quick Verification R&S FSH
1145.5973.82 2.4 E-11
Verify the tracking generator output level
Test principle The generator output must be connected to the RF input and verified.
Test equipment None
Test setup Connect the generator output of the R&S FSH to the RF input.
R&S FSH settings
- [ PRESET ]
- [ MEAS : TRACKING GEN :]
Verification
Verify that the level of the sweep shows 0 dB +5 /-7 dB.
Following internal calibration, verify that the level shows 0 +/-1 dB.
Adjustment functions
Caution:
Only qualified personnel should carry out the re-alignment since any change
substantially influences the measurement accuracy of the instrument. For this
reason, the calibration program is available only on the R&S Service Board.