Module Replacement R&S FSH
1145.5973.82 3.12 E-11
Replacing the housing
Housing parts:
2.1: Bottom cover
2.2: Rear case
2.3: Grip cover
2
.4: Top cover
Notes: When replacing the rear case, the existing type plate must be placed on the new rear
case, or the old series number must be written on the new type plate.
When the grip cover is replaced, the “input unit power/audio connections” have to be
placed in the new grip cover. See also Replacing the power/audio connections.
Open the instrument.
Replace the specific housing part.
Close the instrument.
Perform the quick verification test.
Refer to Chapter 2 “Adjustment”.
Type Plate
2.1
2.2
2.3
2.4
Fig. 3-7 Replacing the housing