GUI Reference
R&S
®
ZNB/ZNBT
282User Manual 1173.9163.02 ─ 53
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "DC1d1" |
"DC2d1" | ...
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "DC1d1" |
"DC2d1" | ...
Ranges
"DC 1", ..., "DC 4" in the "Ranges" section configure the measurement ranges of DC
INPUT 1, ..., 4, respectively. For best accuracy, adjust the ranges to the measured sig-
nals. To protect the instrument, all ranges are pre-set to "+/- 20 V" (the widest range).
If the actual voltage of the input signal exceeds the selected range, the measurement
is incorrect, and a DC overload message is displayed.
Remote command:
[SENSe<Ch>:]DC<DCInp>:RANGe
6.2.11.2 Power Added Efficiency (PAE) Dialog
The "Power Added Efficiency (PAE)" dialog configures the measurement of the PAE of
an active 2-port device.
Access: TRACE – [MEAS] > "DC" > "PAE..."
Background information
Refer to section Chapter 5.7.18.1, "Power Added Efficiency", on page 234.
Meas Softtool