ACQuire:PEAKdetect <PeakDetect>
Enables or disables the peak detect acquisition mode.
You can use this command alternatively to CHANnel<m>:TYPE.
Parameters:
<PeakDetect> AUTO | OFF
*RST: OFF
ACQuire:HRESolution
<HighRes>
Enables or disables the high resolution acquisition mode.
You can use this command alternatively to CHANnel<m>:TYPE.
Parameters:
<HighRes> AUTO | OFF
*RST: OFF
ACQuire:NSINgle:COUNt <NSingleCount>
Sets the number of waveforms acquired with RUNSingle.
Parameters:
<NSingleCount> Number of waveforms.
Range: 1 to maximum number that depends on the record
length.
*RST: 1
ACQuire:AVERage:COUNt
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
Parameters:
<AverageCount> Range: 2 to 100,000
*RST: 2
ACQuire:AVERage:RESet
Deletes the waveform and restarts the average calculation.
Usage:
Event
ACQuire:AVERage:COMPlete?
Returns the state of averaging.
Return values:
<AverageComplete> 0 | 1