EasyManua.ls Logo

Rohde & Schwarz RTA4000 - Page 579

Rohde & Schwarz RTA4000
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTA4000
579User Manual 1335.7898.02 ─ 08
CHANnel<m>:DATA:XORigin?
CHANnel<m>:DATA:ENVelope:XORigin?
CALCulate:MATH<m>:DATA:XORigin?
MASK:DATA:XORigin?
LOGic<p>:DATA:XORigin?
DIGital<m>:DATA:XORigin?
REFCurve<m>:DATA:XORigin?
Return the time of the first sample of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Xorigin> Time in s
Example: See Chapter 17.2.1.4, "Reading waveform data in unsigned inte-
ger format", on page 426
Usage: Query only
CHANnel<m>:DATA:XINCrement?
CHANnel<m>:DATA:ENVelope:XINCrement?
CALCulate:MATH<m>:DATA:XINCrement?
MASK:DATA:XINCrement?
LOGic<p>:DATA:XINCrement?
DIGital<m>:DATA:XINCrement?
REFCurve<m>:DATA:XINCrement?
Return the time difference between two adjacent samples of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Xincrement> Time in s
Example: See Chapter 17.2.1.4, "Reading waveform data in unsigned inte-
ger format", on page 426
Usage: Query only
CHANnel<m>:DATA:YORigin?
CHANnel<m>:DATA:ENVelope:YORigin?
CALCulate:MATH<m>:DATA:YORigin?
MASK:DATA:YORigin?
LOGic<p>:DATA:YORigin?
Documenting results

Table of Contents

Other manuals for Rohde & Schwarz RTA4000

Related product manuals