Remote control commands
R&S
®
RTA4000
631User Manual 1335.7898.02 ─ 08
Suffix:
<b>
.
1..4
Parameters:
<Polarity> POSitive | NEGative
POSitive = high active
NEGative = low active
*RST: POS
BUS<b>:UART:SSIZe <SymbolSize>
Sets the number of data bits in a message.
Suffix:
<b>
.
1..4
Parameters:
<SymbolSize> Range: 5 to 9
Increment: 1
*RST: 8
Default unit: Bit
BUS<b>:UART:PARity <Parity>
Defines the optional parity bit that is used for error detection.
Suffix:
<b>
.
1..2
Selects the bus.
Note: SPI and UART protocols occupy two bus lines.
Parameters:
<Parity> ODD | EVEN | NONE
*RST: NONE
BUS<b>:UART:SBIT <StopBitNumber>
Sets the stop bits.
Suffix:
<b>
.
1..2
Selects the bus.
Note: SPI and UART protocols occupy two bus lines.
Parameters:
<StopBitNumber> B1 | B1_5 | B2
1 stop bit, 1.5 stop bits or 2 stop bits are possible.
*RST: B1