Remote control commands
R&S
®
RTA4000
635User Manual 1335.7898.02 ─ 08
Return values:
<FrameCount> Total number of decoded frames.
Usage: Query only
BUS<b>:UART:FRAMe<n>:STARt?
BUS<b>:UART:RX:FRAMe<n>:STARt?
BUS<b>:UART:TX:FRAMe<n>:STARt?
Return the start time of the specified frame.
Suffix:
<b>
.
1..4
<n> *
Selects the frame.
Return values:
<TxFrameStart> Time in s, range depends on sample rate, record length, and
time base
Usage: Query only
BUS<b>:UART:FRAMe<n>:STOP?
BUS<b>:UART:RX:FRAMe<n>:STOP?
BUS<b>:UART:TX:FRAMe<n>:STOP?
Return the end time of the specified frame.
Suffix:
<b>
.
1..4
<n> *
Selects the frame.
Return values:
<TxFrameStop> Time in s, range depends on sample rate, record length, and
time base
Usage: Query only
BUS<b>:UART:FRAMe<n>:STATe?
BUS<b>:UART:RX:FRAMe<n>:STATe?
BUS<b>:UART:TX:FRAMe<n>:STATe?
Returns the status of the specified frame.
Suffix:
<b>
.
1..4
<n> *
Selects the frame.
Return values:
<TxFrameState> STER | SPER | PRER | BRE | OK | INS
Serial bus analysis