EasyManua.ls Logo

Rohde & Schwarz RTE Series - Displaying the Waveform History

Rohde & Schwarz RTE Series
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trying Out the Instrument
R&S
®
RTE
48Getting Started 1326.1026.02 ─ 08
3. Touch and hold the waveform and move your finger slowly in horizontal direc-
tion.
T
ip: You can turn the NAVIGATION knob to change the zoom factor while
holding the waveform.
4. Release the finger when the waveform segment of interest is visible in the
zoom.
5. Tap "Keep" to convert the fingertip zoom into a standard zoom diagram.
5.5 Displaying the W
aveform History
During a continuous acquisition, the instrument stores the acquired data in the
memory and shows the current acquisition on the display
. The history accesses
and displays the samples that were saved before the current acquisition. When
the acquisition was stopped and a new acquisition is started with [RUN CONT] or
[RUN N× SINGLE], the memory is cleared and written anew.
In the following example, you acquire 10 waveforms, and display the 3 most
recent waveforms.
Displaying the Waveform History
www.allice.de
Allice Messtechnik GmbH

Other manuals for Rohde & Schwarz RTE Series

Related product manuals