Voltage Sensors ¸URV5-Z 11
Zero offset and display noise (relevant with small voltages/powers only)
DC probe: zero offset ρ0.5 mV, display noise negligible. All other probes: see diagram. For display noise of the Level Meter ¸URV35 see relevant specifications.
Temperature effect
1)
RF Probe ¸URV5-Z7
without/with accessories
Insertion Units
¸URV5-Z2/-Z4
DC Probe
¸URV5-Z1
18θC to 28θC included in measurement uncertainty
10θC to 40θC 0.17 dB (2 %) 0.17 dB (2 %)
0.02 dB (0.25 %)
0θC to 50θC 0.44 dB (5 %)
0.44 dB (5 %) 0.04 dB (0.5 %)
1)
Residual uncertainty (approx. two standard deviations) after internal temperature correction taking into account the temperature characteristic of the sensor and its temperature. The specified temperature
effect does not apply to the frequency ranges specified in footnote 11 (page 10).
Black curve:
Zero offset (within 1 h after
zero adjustment; temperature
variation <1θC; meter with
probe run in for about 2 hours)
Blue curves:
Display noise (2 standard devia-
tions, monitoring time 1 min,
probe temperature 18°C#to
28θC, values about double at
0θC)
Filter designations:
0 to 11 for the ¸URV55,
¸NRVS and ¸NRVD,
F0 to#F5 for the ¸URV5 and
¸NRV
Measured value
Zero offset (±)
Noise (2σ)
1 dB
11%
0.1 dB
1.1%
0.01 dB
0.11%
0.001 dB
0.01%
0/F5
3/F4
5/F311/F0 9/F1 7/F2
All level and power values refer to 50 :. Values for 75 : differ slightly.
RF probe without plug-on divider 200 ΠV 500 ΠV 1 mV 2 mV 5 mV 10 mV
RF probe in 50 :/75 :#adapter &60 dBm &53 dBm &47 dBm &40 dBm &33 dBm
&
27 dBm
10 V Insertion Unit ¸URV5-Z2 1 nW 5 nW 20 nW 100 nW 500 nW 2 ΠW
RF probe with 20 dB plug-on divider 2 mV 5 mV 10 mV 20 mV 50 mV 100 mV
100 V Insertion Unit ¸URV5-Z4 &40 dBm &33 dBm &27 dBm &20 dBm &13 dBm &7 dBm
100 nW 500 nW 2 ΠW 10 ΠW 50 ΠW 200 ΠW
RF probe with 40 dB plug-on divider 20 mV 50 mV 100 mV 200 mV 500 mV 1 V
&20 dBm &13 dBm &7 dBm 0 dBm +7 dBm
+13 dBm
10 ΠW 50 ΠW 200 ΠW 1 mW 5 mW 20 mW
negligible.