Preparing for Use
R&S
®
ZNLE
15Getting Started 1323.2873.02 ─ 08
2. NOTICE! Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S ZNLE in a rack
1. Use an adapter kit to prepare the R&S ZNLE for rack mounting.
a) Order the rack adapter kit designed for the R&S ZNLE. For the order num-
ber, see the data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S ZNLE to shelf height.
3. Grab the handles and push the R&S ZNLE onto the shelf until the rack brack-
ets fit closely to the rack.
4. Tighten all screws in the rack brackets with a tightening torque of 1.2 Nm to
secure the R&S ZNLE in the rack.
To unmount the R&S ZNLE from a rack
1. Loosen the screws at the rack brackets.
2. Remove the R&S ZNLE from the rack.
3. If placing the R&S ZNLE on a bench top again, unmount the adapter kit from
the R&S ZNLE. Follow the instructions provided with the adapter kit.
3.5 Connecting to Power
The R&S ZNLE is equipped with an AC power supply connector.
3.5.1 Connecting the AC Power
The R&S ZNLE can be used with different AC power voltages and adapts itself
automatically to it. Refer to the datasheet for the requirements of voltage and fre-
quency.
For safety information, see "Connecting to power" on page 7.