17
E1
Error Code Cause
· Handpiece Overload
· Operation as condition
open the chuck
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minutes
at the cool place.
· If the problem is not resolved, contact the manufacturer.
· Use after closing the chuck of handpiece.
· Defective handpiece assembly
· Defective bearing
· Defective motor
· Contact the manufacturer.
Remedy
E0
Error Code Cause Remedy
Error Code
- FORTE 100
- FORTE 100α, FORTE 100αD
E1
· Defective connection of
handpiece
· Abnormal operation of motor
· Handpiece Overload
· Operation as condition
open the chuck
· Check the connect condition of power cord and
handpiece cord.
· If the problem is not resolved, contact the manufacturer.
· Re-turn on the power.
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minute
s
at the cool place.
· f the problem is not resolved, contact the manufacturer.
· Use after closing the chuck of handpiece.
· Defective handpiece assembly
· Defective bearing
· Defective motor
· Contact the manufacturer.