"Hi" High level reset Beeps are made to reset the high level by
pressing forward/reverse selector (Rev.)
switch while completely depressing knee
control panel .
"Lo" Low level reset Beeps are made to reset the low level by
pressing forward/reverse selector (Rev.) switch
, without depressing knee control panel .
"rc" Recovery
The magnet sensor reset mode is returned to
the main mode by pushing the
forward/reverse selector (Rev.) switch.
Reset mode Function How to reset
9.3 Magnet sensor reset mode
Beeps are made 2 seconds after turning on the power switch while pressing both turbo switch
and forward/reverse selector(Rev.) switch at a time, which mean the reset to magnet
sensor mode. BY turning speed control knob , it is possible to reset the mode to “Hi,” “Lo”
and “rc” modes.
CAUTION
For safety, the above conditions have been already set when shipped and, therefore, do not
change them.
9.4 Specifications
9.4.1 Control unit
- Model: SDE-S60/L60
- Power supply: AC200~240V, 50/60Hz
- Weight: 2.9kg
- Dimension: 94mm(W) x 268mm(D) x 228mm(H)
9.4.2 Motor handpiece
- Model: SDE-BH60S
- Speed: 1,000~50,000RPM
- Weight: 230g(not including the motor cord)
- Dimension: L159 x Ø27
- Cord length: 1.4m
9.4.3 Foot pedal
- Model: SDE-FS60
- Weight: 384g
- Cord length: 1.8m
9.4.4 Control unit stand
- Weight: 152g
- Dimension: 122mm(W) x 268mm(D) x 30mm(H)
9.4.5 Motor handpiece stand
- Weight: 68g
- Dimension: 60mm(W) x 110mm(D) x 45mm(H)