Do you have a question about the Salter Labs 7100 and is the answer not in the manual?
Prepare humidifier, add sterile distilled water to maximum level, and ensure lid is secure.
Attach the bubble humidifier to the oxygen source, like a flowmeter or concentrator.
Test for leaks by blocking outlet port at 6 LPM; an alarm should sound within 5 seconds.
Attach oxygen tubing, adjust flow, apply interface, and refill water as needed.
Wash the bubble humidifier with warm soapy water every 14 days.
Rinse thoroughly with distilled or boiled water and replace the humidifier every 30 days.
If using tap water, wash the humidifier every 3 days.