EasyManua.ls Logo

Salter Labs 7900 - User Manual

Salter Labs 7900
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
Home
Use
7900 Bubble Humidifier Technical Bulletin
7900 High Flow Bubble Humidifier- 6 PSI Safety Pop-off- 6 LPM to 15 LPM
The 7900 High Flow Bubble Humidifier is a disposable, single-patient
use device for use in health care facilities and home care. The
humidifier is used to provide additional humidity to oxygen therapy for
flow rates at 6 to 15 liters per minute (LPM). The Bubble Humidifier
has a 6 PSI safety pop-off valve, which sounds and releases excess
pressure if the oxygen tubing becomes blocked.
Product Use
1. Wash your hands and remove the Bubble Humidifier from
packaging.
2. Use sterile, distilled or boiled tap water. If using boiled tap water,
cool water to room temperature before adding to the Bubble
Humidifier.
3. Unscrew the lid and add water to the maximum level marking
(290 ml) located on the side of the bottle.
4. Replace the lid. Ensure that the threads are properly aligned and
that the lid is screwed on tight.
5. Attach the Bubble Humidifier to an oxygen source, e.g. flowmeter,
oxygen concentrator or liquid tank.
6. Test for leaks. Turn the flow to 6 LPM and block the oxygen outlet
port on the Bubble Humidifier (A). A sound should emit within
5 seconds. If there is no sound, check the lid and other connections
for leaks. Repeat the leak test.
7. Attach the oxygen supply tubing to the oxygen outlet port (B) and
adjust oxygen flow to the prescribed setting.
8. Check for bubbles in the water. If there are no bubbles check for
leaks. If the device sounds, check for tubing obstruction.
9. Apply your high flow nasal cannula or oxygen mask.
10. If water level drops below the minimum level marking on the
Bubble Humidifier add more water.
11. Empty and replace the water in Bubble Humidifier daily.
(Continue on the next page.)
SLML 112 Rev C 03/2019
Oxygen
Tubing
To
Patient
Oxygen Source
(B)
Oxygen Source
Pressure
Relief
Valve
Diffuser
Oxygen
Outlet
Port
Minimum
Water level
Maximum
Water Level
Oxygen Source
(A)
Question and Answer IconNeed help?

Do you have a question about the Salter Labs 7900 and is the answer not in the manual?

Summary

Product Use

Water Preparation and Filling

Instructions on using sterile, distilled, or boiled tap water and filling the humidifier to the maximum level.

Leak Testing and Attachment

Steps for attaching the humidifier, testing for leaks by flow rate, and checking for bubbles.

Daily Maintenance and Replacement

Guidance on daily water replacement and applying the nasal cannula or oxygen mask.

Cleaning Instructions

Washing Procedure

Details on washing the humidifier with soapy water, frequency for tap water use, and avoiding harsh cleaners.

Rinsing and Inspection

Instructions for rinsing the device thoroughly and visually inspecting it for cleanliness.

Replacement Schedule

Guidance on replacing the Bubble Humidifier every 90 days or sooner if mineral deposits persist.

Salter Labs 7900 Specifications