Document Reference: CLIN-UMAN-002426 Revision: 13.00 Page 1 of 48
Evoke™ SCS System User Manual
Evoke SCS System Instructions for Use

Table of Contents

Other manuals for SALUDA MEDICAL Evoke

Questions and Answers

  • P
    Philip OconnellSep 15, 2025
    What does it mean if the Contact Clinician Indicator Light is on either the Charger or EPC for SALUDA MEDICAL Evoke?
    • D
      Daniel SmithSep 15, 2025
      If the Contact Clinician Indicator Light is on either the Charger or EPC, the stimulator is in Safe Mode. Note down the error on the screen and perform a ‘Soft Reset and Reconnect’.
  • S
    Sean GardnerAug 25, 2025
    Why SALUDA MEDICAL Evoke Charger will not turn on?
    • C
      Clayton MasonAug 25, 2025
      If your SALUDA MEDICAL Medical Equipment Charger won't turn on, or if it beeps three times and turns off during charging, press and hold the ‘Start Charging’ button for 2 seconds. If nothing happens or the Charger battery indicator shows below 1 bar, plug the Charger into the power adapter. Once fully charged, disconnect from the power adapter and try charging again.
  • K
    Kimberly BerryAug 26, 2025
    Why does my SALUDA MEDICAL Evoke Charger beep continuously during charging?
    • J
      Jennifer OrtizAug 26, 2025
      If your SALUDA MEDICAL Medical Equipment Charger beeps continuously while charging, try repositioning the Charger coil over the implant. Ensure the ‘Charging Link’ indicator displays at least 2 bars.
  • T
    Timothy SloanAug 26, 2025
    What to do if my SALUDA MEDICAL Evoke Charger started beeping during charging?
    • C
      Cathy GrossAug 27, 2025
      If your SALUDA MEDICAL Medical Equipment Charger starts beeping during charging, adjust the coil position until the ‘Charging Link’ indicator shows at least two bars.
  • S
    Scott JohnsonSep 16, 2025
    What to do if I see 'Current at max/Control variable at max' on my SALUDA MEDICAL Medical Equipment?
    • J
      Jonathan CrawfordSep 16, 2025
      If 'Current at max/Control variable at max' is displayed: * The maximum current limit may be too low. If the patient does not report any surges in stimulation and only cutting out, increase the maximum current limit. * There may be high artefact. Check the patient sensitivity curve; if it is not the correct shape, try changing recording or reference electrodes. Increase pulse width to reduce current requirement. * There may be an electrode disconnection. Check the impedances. * Stimulation is on in closed loop mode, and the charge coil is close to the stimulator whilst charging has been started via the Battery Charger start button. Turn off the Charger and see if the problem persists.
  • Z
    Zachary ThompsonSep 23, 2025
    What to do if my SALUDA MEDICAL Evoke Medical Equipment electrode is disconnected?
    • E
      Erica ConnerSep 23, 2025
      If an electrode is disconnected: * The electrode may be broken or in highly resistive tissue; reprogram using a different electrode. * The electrode may be disconnected from the lead adapter or stimulator; remove the proximal connector end from the lead adapter (or CLS if permanent implant procedure), clean the contacts, and re-insert.
  • M
    Mr. Charles Johnson PhDSep 7, 2025
    What to do if telemetry connection fails due to time out on SALUDA MEDICAL Evoke Medical Equipment?
    • J
      Jason NguyenSep 7, 2025
      If the telemetry connection fails due to a time out, the CST may be too far from the stimulator. Move the CST closer to the stimulator and ensure there is no EMI from other devices.
  • J
    Joseph SmithSep 8, 2025
    What to do if my SALUDA MEDICAL Evoke CI will not turn on?
    • M
      Matthew TorresSep 8, 2025
      If your CI will not turn on, the CI battery may be flat. Attach the power supply to the CI and the power outlet, ensuring the power is turned on.
  • U
    uriveraSep 2, 2025
    What to do if my SALUDA MEDICAL Equipment CI does not respond or is slow to respond?
    • W
      William FischerSep 2, 2025
      If your SALUDA MEDICAL Equipment CI is not responding or is slow to respond, this may be because the CI operating system may not be functioning correctly. Restart the CI by shutting down or restarting it. If the CI does not respond, shut down the CI by pressing the power button for 10 seconds. Then release the power button and press it again to restart the CI.
  • D
    Debbie FordAug 31, 2025
    How to change programs on SALUDA MEDICAL Evoke Medical Equipment?
    • Y
      ywaltersAug 31, 2025
      To change programs on your SALUDA MEDICAL Medical Equipment, press the button once (do not hold it), and then press it again while it is still flushing until you select the program you want.

SALUDA MEDICAL Evoke Specifications

General IconGeneral
ManufacturerSaluda Medical
TechnologyClosed-loop spinal cord stimulation
Battery TypeRechargeable
Device TypeSpinal Cord Stimulator
ProgrammabilityYes
Feedback MechanismEvoked Compound Action Potential (ECAP) monitoring