12
-
HEC-DR5000
(GENERAL)
Error message indication (in Service Mode)
*Abnormal indications detected during operation are stored in a memory up to
seven. Seven abnormal indications stored can be confirmed in the service mode
(Older indications than the seven abnormal indications will be deleted sequentially).
To change over to the service mode, turn ON the POWER switch at the rear of the massager
while simultaneously pressing the OPERATION ON/OFF on a remote controller.
When the "FINE" button is pressed three times with the OPERATION ON/OFF kept pressed,
the massager is placed in the service mode.
(When the massager is placed in the service mode, four blue lamps at the WHOLE BODY
SENSOR AUTOMATIC COURSE come on).
To reset the service mode, turn OFF the POWER switch.
<Resetting Procedure of Abnormal Indication Memory>
Simultaneously press both POSITION ADJUST "UP" and "DOWN"
buttons on the remote controller to delete all abnormal indication
memory (When the seven STIFFNES indicators are all lit up, all
abnormal indication memory have been deleted).
When the massager is placed in the service mode, the latest abnormal indication is displayed.
The POSITION ADJUST " (UP)" button of the remote controller is used to display older
abnormal indications one by one, and " (DOWN)" to display newer abnormal indication one
by one (Following the sixth older (the oldest) indication, the latest indication is displayed again).
Up to seven abnormal indications are stored in a memory. In case of less than seven, since
there are not any abnormal indications, some of the seven STIFFNESS indicators are lit
alternatively to indicate the state that abnormalities are not stored.
POSITION ADJUST "UP" "DOWN" button
"OPERATION ON/OFF" button
"FINE" button (three times)
POWER switch "ON"
All abnormal indication
memory have been deleted
Latest
abnormality
"UP"
"DOWN"
Next older
abnormality
"UP"
"DOWN"
Second older
abnormality
"UP"
"DOWN"
Third older
abnormality
The oldest
abnormality
"UP" "UP"
"DOWN""DOWN"
Fifth older
abnormality
"UP" "UP"
"DOWN" "DOWN"
Fourth older
abnormality
STIFFNESS
RELAX HIGHT
OPEN
LOCK
ON
OFF
POWER