ISO/GLP-compliant Printout/Record
Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series. These parameters include:
GLP header:
Date
Time at beginning of measurement
Balance/scale manufacturer
Balance/scale model
Balance/scale serial number
Software version number
Identification number of the current
sampling operation
GLP footer:
Date
Time at end of measurement
Field for operator signature
The record is output to a Sartorius
data printer or a computer.
52
Settings
Set the following menu codes
(see “Configuring the Balance/Scale”):
ISO/GLP-compliant record
after calibration/adjustment only:
menu code 8 10 2; or
ISO/GLP-compliant record always on:
menu code 8 10 3
Line format for printout: With data ID codes
22 characters: menu code 7 2 2
!No ISO/GLP-compliant record is output if
any of the following settings are configured:
menu codes 6 1 4, 6 1 5, 6 1 6 (automatic
printout) and 7 2 1
Function Keys
Press p to output header and first
measured value.
> Header is output the first time p
is pressed
To output header and reference data
automatically with an application program
active: Press v
End an application:
1) Output GLP footer: Press c
2) End application program: Press c