36
Example: Weighing-in several components
Settings: Application program: Net-total formulation 2 1 6;
Print application parameters: On, print all 7 1 2; Line format for printout:
22 characters 7 2 2; Automatic printout of last net value 7 3 1
Step Key (or instruction) Display/Output
1. Turn on the balance/scale, e
if necessary
2. Place empty container
+ 65.0 g
on the balance/scale.
3. Tare the balance/scale. t
0.0 g
4. Add first component.
+ 120,5 g
5. Store component data. v
0.0 g
NET
COMP1 + 120.5 g
6. Add next component. + 70,5 g
7. Store component data. v
0.0 g
NET
COMP2 + 70.5 g
8. Weigh in further components Repeat steps
if desired. 5 and 6.
9. Fill to target if desired. Press and
+ 191,0 g
hold v (2 sec.) G