Verified Models with EC-type Approval: Specifications
Model ED224S-0CE, ED124S-0CE GK1203-0CE GK703-0CE
ED224S-PCE
Type BD ED 100 BD ED 100 BD ED 100 BD ED 100
Accuracy class
1
) kkkk
Weighing capacity, Max
1
) 220 g 120 g 1200 ct 700 ct
Scale interval d
1
) 0.0001 g 0.0001 g 0.001 ct 0.001 ct
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e
1
) 0.001 g 0.001 g 0.01 ct 0.01 ct
Minimum capacity, Min
1
) 0.01 g 0.01 g 0.1 ct 0.1 ct
Response time (average) s 2.5 2.5 1.5 1.5
Range of use according 0.01 g – 220 g 0.01 g – 120 g 0.1 ct – 0.1 ct – 700 ct
to CD
1
) 1200 ct
Allowable ambient °C +17 to +27 (+63°F to +80°F)
operating temperature
Net weight, approx. kg 4.8 4.8 4.7 4.7
Weighing pan size mm 90 d 90 d 90 d 90 d
Weighing pan size mm 230 230 160 160
Dimensions (W + D + H) mm 230 + 303 + 330 230 + 303 + 260
Model ED623S-0CE ED523S-PCE ED423S-0CE ED323S-0CE ED153-0CE,
ED153-PCE
Type BD ED 200 BD ED 200 BD ED 200 BD ED 200 BD ED 200
Accuracy class
1
) KKKKK
Weighing capacity, Max
1
) g 620 520 420 320 150
Scale interval d
1
) g 0.001 0.001 0.001 0.001 0.001
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e
1
) g 0.01 0.01 0.01 0.01 0.01
Minimum capacity, Min
1
) g 0.02 0.02 0.02 0.02 0.02
Response time (average) s 11111
Range of use according
to CD
1
) g 0.02 – 620 0.02 – 520 0.02 – 420 0.02 – 320 0.02 – 150
Allowable ambient °C +10 to +30 (+50°F to +86°F)
operating temperature
Net weight, approx. kg 3.6
Weighing pan size mm 115 d
Dimensions (W + D + H) mm 230 + 303 + 136
1
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;
applicable to the European Economic Area
54