Display Cause Solution
err 235 Connecting cable not Connect the cable correctly
on model LE26P, connected correctly
LE225D Cable connected to junction Connect the equipment
box on a different balance correctly
The weight readout Unstable ambient conditions Set up the balance
changes constantly (excessive vibration or draft) in another area
at the place of installation
A foreign object is caught Remove the foreign object
between weighing pan and
balance housing
The weight readout The balance was not Calibrate/adjust the balance
is obviously wrong calibrated/adjusted before
weighing
The balance was not Tare or zero the balance
tared/zeroed defore weighing before weighing
If any other errors occur, contact your local Sartorius Service Center!