150
Standard Models
Model ME5 SE2 ME5-F SE2-F
Readability µg 1 0.1 1 0.1
Weighing capacity g 5.1 2.1 5.1 2.1
Tare range (subtractive) g 5.1 2.1 5.1 2.1
Repeatability <±µg 1 0.25 1 0.25
Linearity <±µg 4 0.9 4 0.9
Sensitivity drift within
+10 to +30 °C (50 to 86 °F) <±/K 1·10
–6
1·10
–6
1·10
–6
1·10
–6
Response time (average) s < 10 < 10 < 10* < 10*
External calibration weight g 5 (E2) 2 (E2) 5 (E2) 2 (E2)
(of at least accuracy class...)
Allowable ambient operating temperature +5 ... +40 °C
Operating temperature range +10 ... +30 °C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.1 – 0.4 0.2 – 0.4 0.1 – 0.4 0.2 – 0.4
Pan size mm d 30 d 20 d 50 d 50
Balance dimensions (W + D + H)
Weigh cell mm 122 + 319 + 107 122 + 319 + 107
Display and control unit 122 + 291 + 110 250 + 291 + 110
Weighing chamber height
Weigh cell kg 3.5 3.5 4.3 4.3
Display and control unit 3.5 3.5 3.5 3.5
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection (AC adapter) Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption max. 23 VA
Power consumption, weigh cell and approx. 7 W (typical)
display and control unit only
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per pound,
Chinese taels, momme, Austrian carats, tola, baht, mesghal
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
Specifications