153
Verified Models with EC Type Approval
Model ME5-0CE SE2-0CE
Type KC BN 100 KC BN 100
Accuracy class* kk
Scale interval d* µg 1 0.1
Max. weighing capacity* g 5.1 2.1
Verifcation scale interval e* mg 1 1
Min. capacity* mg 0.1 0.01
Tare range (subtractive) <100% of the max. weighing capacity
Application range* mg 0.1-5100 0.01-2100
Response time (average) s <10
Allowable operating temperature range
1
) 273 … 313 K (0 … +40°C, 32°F … 104°F with “isoCAL” function
External calibration weight g 5 (E2) 2 (E2)
(of at least accuracy class...)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate (depends on filter level) 0.2 – 0.4
Pan size mm d 30 d 20
Balance dimensions (W + D + H)
Weigh cell mm 122 + 316 + 122
Display and control unit 250 + 291 + 110
Weighing chamber height
Weigh cell kg 3.5
Display and control unit 3.5
Net weight, approx. kg 11.1
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 – 60 Hz
Power consumption (AC adapter) max. 23 VA
Power consumption, weigh cell and approx. 7 W (typical)
display and control unit only
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams, carats and milligrams
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
1
) = With the isoCAL function deactivated, the verified balance can be used only within the limited temperature range
(can be modified only by the Sartorius Service Center): For balances of accuracy class k: +15°C to +25°C (59 to 77°F)
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area