26
Internal Calibration/Adjustment 
Default setting:  
SETUP - BAL.SCAL.- CAL.Just. - Cal.Int. 
Voraussetzung:
The weigh cell housing has a built-in motorized calibration/
adjustment weight.
§ Select calibration: Command ESC Z 
>   The internal calibration weight is automatically loaded
>  The balance is calibrated
>   When the setup is configured to “Calibration and adjustment 
in one,” the balance will be adjusted automatically
>   The internal calibration weight is removed
Performing Calibration and Adjustment Routines
The following settings can be configured:
–   Always perform calibration and adjustment in one routine 
(factory setting) 
–   After calibration, the user has the option to quit the routine 
without correction or to adjust the balance.
If no deviations are found during calibration, the calibration/
adjustment routine can be exited after the calibration is completed.  
Two keys are now active:
–  Start the adjustment:   Command ESC f1_
–  Exit the routine:  Command ESC f3_
  Step  Execute interface  Display/ 
    command  Output
1.  Tare balance  ESC T  0.0000 g
2.  Start adjustment routine  ESC Z  CAL.Int.
  The internal calibration weight   CAL.RUN.
  is applied automatically. 
3.  Calibration/adjustment      Cal.end 
  executed
4.   Internal weight is removed    0.0000 g
  from balance
External Calibration
Default setting:  
SETUP - BAL.SCAL.- CAL.Just. - Cal.Ext. 
The required calibration weight is configured at the factory 
(see “Specifications”).
  Step  Execute interface  Display/ 
    command  Output
1.  Tare balance  ESC T       0.0000 g
2.  Start adjustment routine  ESC W  CAL.ext.
   Once you store the zero  
point, a prompt for the    - 50.0000 g 
required calibration weight  
flashes on the display. 
3.  Place displayed calibration       50.0000 g
  weight on balance (in this  
  example: 50 g). 
  Weight too low:    
  a minus sign “-” is shown
  Weight too high:  
  a plus sign “+” is shown
 
   The display stops flashing  
as soon as the weight value  
is within the defined limit. 
4.  Adjustment carried out;    Cal.end 
  adjustment weight is  
  displayed    + 50.0000 g
5.  Remove the adjustment       50.0000 g
  weight