9
Specifications
The climate module internal memory contains parameters for
correcting the raw data from the individual climate sensors
including the associated uncertainty data.
The processing and correction of raw sensor data takes place
in the respective end device to which the climate module has
been adapted (e.g. Cubis mass comparator).
The specific corrected value for the respective climate
parameter (
t, p, hr
) represents the result of the climate
measurement in connection with the associated uncertainty
data
U, k
= 2. The expanded uncertainty
U
is saved in this case
to the climate module memory sensor-specific.
For a calibrated climate module (YCM20MC-DAkkS), the
following standard uncertainties
u, k
= 1 are specified for
the given measurement ranges:
– Temperature:
u
t
= 0.15 K, 18°C <
t
< 27°C
– Air pressure:
u
p
= 1 hPa, 800 hPa <
p
< 1100 hPa
– Humidity:
u
hr
= 1%, 30% <
hr
< 70%
u
hr
= 2%, 70% <
hr
< 90%